Reports and result lists

Overview

All results of QA tests and calibration tasks are stored in QAWeb Enterprise and can be accessed from several different perspectives:

  • For a single workstation the last results are accessible on the Workstation level in the Tasks tab

  • For a single workstation, the history of all results is accessible on the Workstation level, in the History tab

  • Lists and reports spanning multiple workstations are described in this chapter.

All lists have several ways to filter the results using search criteria.

Export to Excel

Most reports and result lists can be downloaded as an Excel file.

To download the current list results as Excel file:
  1. Navigate to a list and optionally filter the data according to your needs

  2. Click the export icon

The maximum number of exported rows is 10,000.

Budget report

The budget report indicates the backlight guarantee status of Barco display models. This information can be helpful in budget decisions by identifying monitors of which the backlight guarantee is close to expiring.

For more details on the Barco warranty terms, please read the Healthcare products warranty policy at our website: Barco.com terms and conditions’)

Task History

The task history list provides an overview of all calibrations and QA tests of all workstations in the organization. The list can be filtered on:

  • Workstation name

  • Display model

  • Display model

  • Serial number

  • Execution date and time

  • Result (Passed / Failed)

  • Display use

JC Report

The Joint Commission report provides an overview of the calibration status of all displays present in your organization.

To open the JC report:
  1. Click QA in the header menu.

  2. In the side menu, in the Reports section, click JC Report.

The following information is present:

  • Workstation name

  • Display model

  • Serial number

  • Display use

  • QA Status

  • Last calibration date and time

  • Calibration target

  • Last luminance response test: result, date and time

QA Test Results

All QA test results can be displayed in an organization-wide lists. For each type of QA test, a result list with different columns will be displayed, depending on the type of test.

To open a test result list:
  1. Click QA in the header menu.

  2. In the side menu, in the Reports section, click Test results.

  3. Select a QA test type.

  4. Optionally, filter and sort the data.

PDF Summary reports

Summary PDF reports combine data of multiple QA test results into one report. The PDF reports can be downloaded for sharing information with persons that do not have a QAWeb Enterprise account, or for archiving purposes.

Note

Summary reports are available for 7 days, after which they are automatically removed from the server. At any time, a new summary report can be generated. A summary report is created using the information that is available at the time of its generation.

The summary report overview lists all recently generated reports. There are two main types of summary reports:

  • Workstation reports contain information of a single workstation in the PDF report.

  • Fleet reports contain information of multiple workstations, and can summarize information on your entire fleet of diagnostic workstations in a single PDF report.

To create a new summary report:
  1. See Viewing and creating summary reports.

To create a new fleet report:
  1. Click QA in the header menu.

  2. In the side menu, in the Reports section, click Summaries.

  3. Click on the Fleet reports tab.

  4. Click on the icon.

  5. Select the type of fleet report to generate. Other parameters may need to be entered, depending on the report type.

The following fleet reports are available:

  • DICOM GSDF fleet report: history of DICOM GSDF calibrations and DICOM GSDF luminance response test results, listed in chronological order.

  • Mammography Compliance fleet report: Contains the most recent mammography acceptance test and a history of mammography constancy test results, listed in chronological order.