Default policies
The default policy is based on ACR recommendations and our manufacturer’s recommendations.
Calibration
Use: Reviewing, Dentistry
Schedule: yearly
Method: default
Black point luminance: according to the display’s recommendation
White point luminance: according to the display’s recommendation
Gamut: display native
Luminance response function: DICOM-GSDF
SteadyColor: standard
Use: General radiology, Mammography, Breast Tomosynthesis
Schedule: half-yearly
Method: default
Black point luminance: according to the display’s recommendation
White point luminance: according to the display’s recommendation
Gamut: display native
Luminance response function: DICOM-GSDF
SteadyColor: standard
Use: Pathology
Schedule: half-yearly
Method: default
Black point luminance: according to the display’s recommendation
White point luminance: according to the display’s recommendation
Gamut: sRGB
Luminance response function: sRGB
SteadyColor: off
DICOM GSDF Luminance Response Test
Use: General radiology
Schedule: quarterly
Error tolerance: <15%
Sensor: internal
Use: Dentistry
Schedule: yearly
Error tolerance: <15%
Sensor: internal
sRGB Luminance Response Test
Use: Pathology
Schedule: Monthly
Error tolerance: < 10%
Sensor: internal
Grayscale chromaticity test
Use: Pathology
Schedule: quarterly
Delta u’v’: <0.01
Sensor: internal
ICC Compliance test
Use: Pathology
Schedule: Half-yearly
Average dE2000 <= 2
Maximum dE2000 <= 5
Sensor: internal
Luminance ratio test
Use: General radiology, Dentistry
Schedule: Quarterly
Luminance ratio: >= 250
Sensor: internal
Mammography compliance test
Use: Mammography, Breast Tomosynthesis
Schedule constancy test: Weekly
White point luminance: < 5% relative to the latest calibration
Error tolerance luminance response: < 10%
Multi-display uniformity: < 10%
Sensor: internal
Maximum luminance test
Use: Reviewing
Schedule: Yearly
Luminance: >= 150 cd/m²
Sensor: internal
Use: General radiology, Pathology, Dentistry
Schedule: Quarterly
Luminance: >= 350 cd/m²
Sensor: internal
SteadyColor response test
Use: Mammography, Breast Tomosynthesis
Schedule: Half-yearly
Error tolerance: <15%
White point chromaticity test
Use: Pathology
Schedule: Quarterly
Delta u’v’: < 0.01
White point chromaticity: D65
Sensor: internal
sRGB Luminance Response Test
Use: Pathology
Schedule: Monthly
Error tolerance: < 10%
Sensor: internal
sRGB Color Space Test
Use: Pathology
Schedule: Half-yearly
Average dE2000 <= 2
Maximum dE2000 <= 5
Sensor: internal
By default, QAWeb Enterprise uses a room ambient light of 35 cd/m².